Method design
Keys, CalGroup, Origin/blank corr, Method Validation, RecPct and limits.
Keys, import modes, ESTD/ISTD, doses, area thresholds, and calibration mode — the method tables you edit before Analyze.
Keys
| Key | Meaning |
|---|---|
| AcqKey | Acquisition / method set. Append import creates a new AcqKey so several instruments or acquisition methods share one Quanto method. |
| QntKey | Quant ion / transition identity within an analyte. |
| SmpKey | Sample key for grouping injections (Key Review winners). |
Multi-instrument pick: Key Review.
Replace vs Append (method import)
Method → Load / Import:
| Mode | Effect |
|---|---|
| Replace | Swaps the current method; sample results are cleared |
| Append | Adds the import under a new AcqKey; keeps existing analytes |
You can append only selected AcqKey + name keys when the dialog offers a candidate list. Details: Reports, import, and export.
ESTD vs ISTD
| Setting | Meaning |
|---|---|
| ESTD | Quant has no ISTD link — external standard curve |
| ISTD | Quant points at an ISTD Quant — ratio / internal standard |
You can mix ESTD and ISTD Quants in one method. Guides: ESTD, ISTD. Curve Origin can be Ignore, Include, Force, or Blank offset (blank response as intercept). Blank concentration correction is a separate Quant checkbox that subtracts blank concentration toward NativeConc.
Doses
Shared dose names on the method:
| Dose | Typical use |
|---|---|
| C1, C2, … | Calibration levels |
| IS | Reserved ISTD spike amount |
Calibration bottles need a dose that exists on the analyte. ISTD bottles still use IS. Partial update of stored levels: Multi-batch dose responses.
Area thresholds
When importing integrator peaks, insignificant peaks are also filtered by a method/channel area (or height) threshold:
| Setting | Default |
|---|---|
| Area threshold percent (method default) | 30 |
| Dose reference | Lowest calibration concentration (or a named dose) |
Threshold = calibration response × percent / 100. Channels can override with an absolute threshold instead of the general percent. See Manual integration.
CalGroup
CalGroup is a text label on bottles that share one calibration story. Unlabeled bottles share the default group.
| Mode on the group | Meaning |
|---|---|
| Calibration | Normal external/internal standard curve for that group |
| StdAdd | Standard-addition line for that group |
Details: CalGroup & standard addition. Blank offset / blank concentration Average (CalGroup) modes stay inside the group — see Blank offset and Blank concentration.
Curve origin and blank corrections
| Setting | What it does |
|---|---|
| Origin = Blank offset | Blank response becomes the curve intercept (b) |
| Blank corr (blank concentration) | Subtracts blank concentration toward NativeConc after the curve is read |
If both are set, Blank offset wins for the curve — blank concentration still applies to NativeConc afterward. Guides: Blank offset, Blank concentration.
Calibration mode
| Mode | Behavior |
|---|---|
| Initial | One curve from the calibration set (default for new work) |
| Bracket | A curve per calibration bracket around samples |
Vendor method import does not set Calibration mode — you choose it in Quanto. Guide: Bracket calibration. CalGroup is separate from Initial vs Bracket: groups split which bottles share a curve; Initial/Bracket decide how cal sets are cut in time.
Method Validation view
After Analyze, the Method Validation pane summarizes curve quality, recovery, and limit flags for the current method/batch. Use it when you are checking whether the method (not only one sample) is fit for purpose. Guide: Method Validation.
Recovery and result limits
Known-dose bottles (Calibration, QC, and any bottle with a dose you treat as known) get:
| Field | Meaning |
|---|---|
| RecPct | Recovery % = 100 × calculated concentration / dose |
| DevPct | RecPct − 100 |
| RecFlag | Pass/fail against recovery limits |
Other limits you can set on the method / Quant include TooLow (usually LOD or Trace), Low (usually LOQ), High, MinCurveR2, MinRsp / MaxRsp, MinQValue, qualifier counts, MinCoElution, and AboveCurveMaximum / PassFailFlag. Plain-language guide: Result limits & flags.
Checklist before Analyze
- Analytes / Quants / Qualifiers and extraction windows.
- AcqKeys after any Append.
- Doses C1… and IS; Quant doses on the curve.
- Sample types mapping.
- ESTD/ISTD links, and height instead of area if you quantitate by height.
- Area threshold percent / dose.
- Calibration mode Initial vs Bracket; CalGroup labels / StdAdd where needed.
- Origin / Blank corr / Blank use if you use blanks.
- Result limits (R², response, Q-value, recovery) you care about.
Then load bottles and Analyze (Set up an analysis).