Result limits & flags
RecPct / DevPct / RecFlag, MinCurveR2, response and qualifier limits.
Limits tell Quanto when a result or curve point is outside what you accept. After Analyze, flags show up on results and in Method Validation.
Recovery on known doses
When a bottle has a known dose (Calibration, QC, and any other bottle you treat as known):
| Field | Formula / meaning |
|---|---|
| RecPct | 100 × calculated concentration / dose |
| DevPct | RecPct − 100 |
| RecFlag | Pass/fail against the recovery limits you set |
There is no separate Spike sample type in the bottle list — put the known amount on Dose (for standard addition, use the spike increment). See Set up an analysis and CalGroup & standard addition.
ISTD and ESTD curves both produce concentrations that feed RecPct the same way once a dose is present.
Curve and response limits
| Limit | Typical use |
|---|---|
| MinCurveR2 | Minimum acceptable coefficient of determination for the fit |
| MinRsp / MaxRsp | Response (area or height) window for a peak |
| MinQValue | Minimum qualifier / confirmation value |
| Qualifier counts | How many qualifiers must pass |
| MinCoElution | Minimum co-elution / overlap score when the method uses it |
| AboveCurveMaximum | Flag when the result sits above the top calibrator |
| PassFailFlag | Overall pass/fail roll-up for the result |
Set these on the method / Quant tables before Analyze. Names match the columns you see in Quanto; leave a limit empty when you do not want that check.
Concentration limits
Set TooLow, Low, and High on Quant, Limit-group limits, or the Requested matrix. Headers are TooLow and Low; tooltips say TooLow is usually LOD or Trace, and Low is usually LOQ. FastPositive, carry-over, and concentration flags use Low (and TooLow where the detection-vs-quant band matters). There is no separate reporting-limit (RL) field.
Unknown-sample results can show a text form of final concentration (Results grid FConc text, Key Review SuperCell, reports via Result.FinalConcentrationText):
| Situation | Display |
|---|---|
| Method No-peak text is set (Properties → Results), there is no usable peak, and there is no FConc override | That text (example: n.d.) |
| Method TooHigh prefix is set and the result is TooHigh / at or above High | Prefix + formatted High (example: > 10) |
| Method Below-Low prefix is set and reported concentration is below Low | Prefix + formatted Low (example: < 1) |
| Otherwise (empty option, or none of the cases above) | The usual numeric final concentration |
Numeric FConc / Result.FinalConcentration in reports and Key Review sums is the effective reported value: a finite Results FConc override when the analyst typed one, otherwise Native × Dilution. This is not the five-level ConcentrationFlag (TooLow / Low / None / High / TooHigh).
User override of final concentration
Results FConc is editable. Type a number to store Result.ManualFinalConcentration (batch TSV on the result row). Clear the cell — or Key Review context menu Clear FConc override — to restore computed Native × Dilution. Analyze refreshes the computed value and does not clear the override.
When an override is set:
- Reports (
Result.FinalConcentration/FinalConcentrationText), Key Review SuperCell text, Amount / group numeric sums, and the Concentration flow FinalConc card use the override. - FConc text skips no-peak text (the analyst supplied a number); TooHigh / below-Low prefixes can still apply.
- Sample-key auto ranking still uses computed Native × Dilution so an override does not change which SuperCell is auto-picked.
- The Results FConc cell is tinted; Key Review flags include Override; the flow FinalConc caption reads Manual override.
Where flags appear
- Results grid — per sample / Quant
- Method Validation pane — method-wide scan
- Reports — include flag / RecPct columns with Q. functions when your workbook needs them (Report formulas)
Practical tips
- Start with recovery limits on QC (and maybe Calibration) before tightening R² and response windows.
- After clearing a bad cal point (Use off), Analyze again so RecPct and curve flags refresh.
- Blank offset and blank concentration change concentration — re-check RecPct afterward (Blank offset, Blank concentration).
- For ISTD work, also watch ISTD response limits so a dead ISTD does not silently pass targets (ISTD).