Limits tell Quanto when a result or curve point is outside what you accept. After Analyze, flags show up on results and in Method Validation.

Recovery on known doses

When a bottle has a known dose (Calibration, QC, and any other bottle you treat as known):

Field Formula / meaning
RecPct 100 × calculated concentration / dose
DevPct RecPct − 100
RecFlag Pass/fail against the recovery limits you set

There is no separate Spike sample type in the bottle list — put the known amount on Dose (for standard addition, use the spike increment). See Set up an analysis and CalGroup & standard addition.

ISTD and ESTD curves both produce concentrations that feed RecPct the same way once a dose is present.

Curve and response limits

Limit Typical use
MinCurveR2 Minimum acceptable coefficient of determination for the fit
MinRsp / MaxRsp Response (area or height) window for a peak
MinQValue Minimum qualifier / confirmation value
Qualifier counts How many qualifiers must pass
MinCoElution Minimum co-elution / overlap score when the method uses it
AboveCurveMaximum Flag when the result sits above the top calibrator
PassFailFlag Overall pass/fail roll-up for the result

Set these on the method / Quant tables before Analyze. Names match the columns you see in Quanto; leave a limit empty when you do not want that check.

Concentration limits

Set TooLow, Low, and High on Quant, Limit-group limits, or the Requested matrix. Headers are TooLow and Low; tooltips say TooLow is usually LOD or Trace, and Low is usually LOQ. FastPositive, carry-over, and concentration flags use Low (and TooLow where the detection-vs-quant band matters). There is no separate reporting-limit (RL) field.

Unknown-sample results can show a text form of final concentration (Results grid FConc text, Key Review SuperCell, reports via Result.FinalConcentrationText):

Situation Display
Method No-peak text is set (Properties → Results), there is no usable peak, and there is no FConc override That text (example: n.d.)
Method TooHigh prefix is set and the result is TooHigh / at or above High Prefix + formatted High (example: > 10)
Method Below-Low prefix is set and reported concentration is below Low Prefix + formatted Low (example: < 1)
Otherwise (empty option, or none of the cases above) The usual numeric final concentration

Numeric FConc / Result.FinalConcentration in reports and Key Review sums is the effective reported value: a finite Results FConc override when the analyst typed one, otherwise Native × Dilution. This is not the five-level ConcentrationFlag (TooLow / Low / None / High / TooHigh).

User override of final concentration

Results FConc is editable. Type a number to store Result.ManualFinalConcentration (batch TSV on the result row). Clear the cell — or Key Review context menu Clear FConc override — to restore computed Native × Dilution. Analyze refreshes the computed value and does not clear the override.

When an override is set:

Where flags appear

Practical tips

  1. Start with recovery limits on QC (and maybe Calibration) before tightening R² and response windows.
  2. After clearing a bad cal point (Use off), Analyze again so RecPct and curve flags refresh.
  3. Blank offset and blank concentration change concentration — re-check RecPct afterward (Blank offset, Blank concentration).
  4. For ISTD work, also watch ISTD response limits so a dead ISTD does not silently pass targets (ISTD).